| VTPC-1 | Gap TPC | VTPC-2 | FTPC 1 | MTPC-L | MTPC-R | FTPC 2 | FTPC 3 | | | |
Wed, 11.03.2026 20:50:00 | 500V/13kV | 500V/8.16kV | 500V/13kV | 500V/17.3kV | 500V/19kV | 500V/19kV | 500V/17.3kV | 500V/17.3kV | Full drift voltage and potential strips ON in big TPCs, 500V on sense wires, Gating Grid turned on. | (BL) | |
Wed, 11.03.2026 21:41:17 | 600V/13kV | 600V/8.16kV | 600V/13kV | 600V/17.3kV | 600V/19kV | 600V/19kV | 600V/17.3kV | 600V/17.3kV | +100V on sense wires, now at 600V | (BL) | |
Wed, 11.03.2026 22:48:02 | 700V/13kV | 700V/8.16kV | 700V/13kV | 700V/17.3kV | 700V/19kV | 700V/19kV | 700V/17.3kV | 700V/17.3kV | +100V on sense wires, now at 700V | (BL) | |
Wed, 11.03.2026 23:48:39 | 800V/13kV | 800V/8.16kV | 790V/13kV | 800V/17.3kV | 800V(SR)/740V(HR)/19kV | 800V(SR)/740V(HR)/19kV | 800V/17.3kV | 800V/17.3kV | +100V except MTPC HR and VTPC2. MTPC and VTPC2 now at 300V below nominal, going to hold there overnight. | (BL) | |
Thu, 12.03.2026 00:07:56 | 900V/13kV | 850V/8.16kV | 790V/13kV | 850V/17.3kV | 800V(SR)/740V(HR)/19kV | 800V(SR)/740V(HR)/19kV | 850V/17.3kV | 850V/17.3kV | +100V on VTPC1, +50V on small TPCs. All TPCs (except VTPC1 sectors 1, 4) are at 300V below nominal | (BL) | |
Thu, 12.03.2026 15:41:21 | 950V(1,4)/900V/13kV | 850V/8.16kV | 790V/13kV | 850V/17.3kV | 800V(SR)/740V(HR)/19kV | 800V(SR)/740V(HR)/19kV | 850V/17.3kV | 850V/17.3kV | +50V on VTPC1 sectors 1, 4. Now all are at 300V below nominal | (BL) | |
Thu, 12.03.2026 23:07:44 | 1050V(1,4)/1000V/13kV | 950V/8.16kV | 890V/13kV | 950V/17.3kV | 900V(SR)/840V(HR)/19kV | 900V(SR)/840V(HR)/19kV | 950V/17.3kV | 950V/17.3kV | +100V. Now all are at 200V below nominal | (BL) | |
Fri, 13.03.2026 16:40:06 | 1100V(1,4)/1050V/13kV | 1000V/8.16kV | 940V/13kV | 1000V/17.3kV | 950V(SR)/890V(HR)/19kV | 950V(SR)/890V(HR)/19kV | 1000V/17.3kV | 1000V/17.3kV | Beginning raise to nominal with +50V to all chambers | (BL) | |
Sat, 14.03.2026 18:46:46 | 1250V(1,4)/1200V/13kV | 1150V/8.16kV | 1090V/13kV | 1150V/17.3kV | 1100V(SR)/1040V(HR)/19kV | 1100V(SR)/1040V(HR)/19kV | 1150V/17.3kV | 1150V/17.3kV | Nominal voltage was reached around 23:00 last night. | (BL) | |
|
Fri, 13.03.2026 17:56:52 At 17:23, began raising GRC voltage as well from its parking value of 1100V, toward nominal 1600V.
Sat, 14.03.2026 18:47:51 During the final raising to nominal, it was discovered that FTPC1 was not actually connected. Had to turn off all FTPC1 voltages plus FTPC2+3 sense wires in order to connect them. FTPC2+3 were raised back to normal relatively quickly, but since FTPC1 had not actually been stressed yet we had to raise it slowly.
Wed, 18.03.2026 07:44:17 In preparation for power cut in EHN1, all TPC voltages and power supplies were turned off.
Wed, 18.03.2026 14:04:54 TPC drift voltage and gating grid were turned on before lunch. Now raising TPCs (and GRC) back to nominal sense wire voltages gradually in steps. |