This is NA61/SHINE electronic logbook (read-only BACKUP!)

Run summary Shift logbook Run settings Beam logbook BPD logbook DAQ logbook DCS logbook Experts logbook Gas logbook TOF logbook Trigger logbook PSD logbook VD logbook Run coordination Detector Upgrade DRS logbook
  DRS logbook, Page 1 of 1  NA61/SHINE logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Subject Text Attachments
  58   Thu, 12.03.2026 15:35:49 D. RybkaNIM splitter in TOF-ROn 2026.03.11 NIM splitter device for trigger
signal was installed in TOF-R rack and connected
to all TOF-R DRS crates.
  
  57   Wed, 25.02.2026 13:36:53 B. MaksiakDRS startup procedureSorry, it's in Polish. When I have more time,
I will translate it and add pictures and
schematics.
  
  56   Thu, 5.02.2026 13:05:28 B. MaksiakNew crates + rearrangementFour new DRS crates (Kontron-Wiener) have
been delivered. They came with their own
power supplies and fan units. At the same
  
  55   Thu, 28.11.2024 20:41:24 D. Rybka, Y. BalkovaDRS board swapped BPD3<->TOF2Thu, 28.11.2024 20:42:01 DRS board 0x30C (BPD3)
changed from SN112 to SN197
Fri, 29.11.2024 12:01:00 DRS board 0x416
  
  54   Mon, 25.11.2024 15:17:25 B. MaksiakBeam counter signals moved from fast to slow board for fragmentation runDue to introduction of A-detector signals,
the signals from S2, S3, V0 i V1 were moved
from fast board (x00f) to slow board (x00e).
  
  53   Mon, 25.11.2024 09:47:49 Aleksandr Dmitrievmove two MRPC boards to TOFFMon, 25.11.2024 09:48:14 AD, DR: We moved
x438 -> 4a0 and x43b -> 4a1
  
  52   Wed, 30.10.2024 18:20:28 B. Maksiak, D. Rybka, Y. BalkovaSwapped names of BMCs of bpd2-02 and 03BMCs (Baseboard Management Controller) of
computers bpd2-02 and bpd-03 had swapped
names.
  
  51   Mon, 2.10.2023 09:20:55 B. MaksiakCrash of psd2 and psd4Mon,  2.10.2023 09:23:06 
Two computers at PSD: psd2 and ps4 crashed
at 2:48 last night. psd4 restarted and lost
  
  50   Mon, 25.09.2023 17:48:16 B. MaksiakSolution for falsely recognized USB portsRecent batch of 4xUSB 3.0 expansion cards
shown a new problem. The system may see all
USB ports, but only some of them may actually
  
  49   Wed, 20.09.2023 12:12:52 B. MaksiakUSB 3.0 card replaced in psd1psd1 computer was falsely showing 4 available
USB ports. Only one port was actually communicating
with the DRS board. The USB card was replaced.
  
  48   Tue, 15.08.2023 15:18:40 B. Maksiakpsd1 turned offDuring data taking the psd1 computer turned
off by itself. This time the computer does
not want to turn back on. PSD was excluded
  
  47   Wed, 9.08.2023 11:49:49 B. MaksiakCasualties during firmware upgrade[LIST]
[*] I accidentally ripped off capacitor C483
from board SN195. It should be checked what
  
  46   Wed, 2.08.2023 19:17:16 B.Maksiak, Y.BalkovaNew DRS firmwareNew firmware revision v6 was prepared for
DRS boards. Its main new features are:
[LIST]
  
  45   Wed, 19.07.2023 14:54:57 B.MaksiakCrash of psd1Last night the machine [I]psd1[/I] crashed
and rebooted. After reboot all USB ports
were visible, but DRS boards could not be
  
  44   Tue, 18.07.2023 16:46:18 B. Maksiak, P. PodlaskiLong busy problem fixedLong busy problem has been fixed.

Two DRS registers (33 and 35) had to be zeroed
 Busy_dobry.pngBusy_zly.png 
  43   Mon, 17.07.2023 17:16:13 B.MaksiakWrong voltage calibration VCalib_2023-07-17_1258.binThe DRS calibration taken about 13:00 had
some issues. [B]Boards x417, x433, x438 in
runs 42292-42297[/B] may be affected by wrong
  
  42   Thu, 13.07.2023 14:18:19 B.MaksiakMajor issue in crate TOF4Yesterday when testing readout I noticed that
all boards in the crate TOF4 do not get eventID
from the trigger system. The LED next to
  
  41   Sat, 1.07.2023 19:23:53 B.MaksiakAnother replacement for USB cards in tof1-01 and tof2-03The boards that had been replaced [URL=http://na61-elog.cern.ch/DRS+logbook/39]few
days ago[/URL] were showing a behavior not
seen before.
  
  40   Wed, 28.06.2023 09:59:35 B.Maksiak, Y.BalkovaDRS boards rearrangements before neutrino runWed, 28.06.2023 09:59:54 
13 DRS boards from BPD1 were removed. Their
USB cables were taken away as well and put
  
  39   Mon, 26.06.2023 18:12:55 B.MaksiakUSB 3.0 cards replaced in tof1-01 and tof2-03tof1-01 and tof2-03 could not show more than
two USB ports. The PCI-Express USB 3.0 expansion
cards were replaced there.
  
  38   Wed, 31.05.2023 14:18:25 B.MaksiakProblems before 2023 neutrino run USB expansion cards on following computers
cannot see 4 USB ports:
[LIST]
  
  37   Wed, 23.11.2022 19:37:16 B. MaksiakBoard x50c (SN142) collected voltage calibration wronglyEarlier this day I took voltage calibration.
During the calibration one of DRS chips of
board x50c did collected voltage calibration
  
  36   Fri, 18.11.2022 14:40:06 B. MaksiakJ110 module in TOF2 replacedAll boards connected to crate TOF2 had channel
9 sine waves missing. The faulty module was
the culprit. It was replaced by new one.
  
  35   Wed, 16.11.2022 11:02:51 B. MaksiakProblems with DRS boards in TOF cratesOn the run meeting 15.11. [URL=https://indico.cern.ch/event/1221341/contributions/5138246/attachments/2547504/4387274/Everything.pdf]I
reported[/URL] bad waveforms in many DRS
boards from TOF crates. There were several
  
  34   Sun, 13.11.2022 16:52:15 B. MaksiakDRS boards rearrangement in TOFOn request of mRPC group, some boards were
moved in between TOF racks. TOF4 supposed
to be populated by 12 boards taken from other
  
  33   Sat, 12.11.2022 22:59:53 B. Maksiakpsd4 computer lost 4 GB of RAMAfter turn off of psd computers yesterday
(11.11.2022 17:30), I had to turn them on
again. psd2 has 18 GB of RAM now. It should
  
  32   Sat, 12.11.2022 16:54:45 B. MaksiakRearrangement of DRS boardsDRS boards configuration was rearranged for
Pb run.
  
  31   Thu, 10.11.2022 21:51:42 B. MaksiakInvestigation of boards 156 and 157This topic is connected with http://na61-elog.cern.ch/DRS+logbook/29

Boards 156 and 157 do not accept event ID
 evid_debug_06.jpgevid_debug_156.jpgevid_debug_157.jpg 
  30   Thu, 10.11.2022 21:27:09 B. MaksiakBoards 142, 143 moved: TOF->PSDThe boards 142 and 143 were moved back (see
http://na61-elog.cern.ch/DRS+logbook/19)
to PSD.
  
  29   Thu, 3.11.2022 14:35:04 B. MaksiakProblems with board SN157The board SN157 shows similar problems as
board [URL=http://na61-elog.cern.ch/DRS+logbook/15]SN156[/URL].
The board does not recognize event ID, produces
  
  28   Wed, 2.11.2022 22:48:25 B. MaksiakProblems with board SN188The board SN188 (problem observed when it
had address x412) has following issues:
[LIST]
  
  27   Tue, 1.11.2022 16:12:29 B. MaksiakDRS computer bpd2-03 needs replacementbpd2-03 machine has one RAM slot malfunctioning.
That was already discovered and labeled last
year, but somehow that machine still got
  
  26   Mon, 19.09.2022 14:03:56 B. MaksiakProblems with board SN151The board SN151 had problems with voltage
calibration. First channel of first DRS chip
seems to not react to changing DAC values
  
  25   Sat, 9.07.2022 11:29:21 B. Maksiakmunmap_chunk, double free or corruption (fixed)During tests related to the previous entry,
six machines (bpd1-0[1-4], bpd2-0[12]) were
giving following errors at the first event
  
  24   Sat, 9.07.2022 11:25:41 B. MaksiakThe readout problems (ctd.)The problem with readout problems reported
in previous entry evolved.
  
  23   Mon, 4.07.2022 15:58:30 B. Maksiak, K. WójcikReadout problems with board x20cThere is a not understandable problem happening
to readout of the board x20c (#C). The board
gets very noisy readout.
 x20c_old_with_signals.png 
  22   Mon, 4.07.2022 15:40:03 B. Maksiak, K. WójcikProblems with boards x207, x209The boards [B]x207[/B] (#06) and [B]x209[/B]
(#09) started to have problems with proper
triggering on events. The board x209 started
 Double_triggering_event_fragments_increase.pngDouble_triggering_excessive_events.png 
  21   Sat, 25.06.2022 13:27:59 B. Maksiak, P. PodlaskiBoard 178 moved. BPD3 -> BPD3The board 178 was moved from serving as readout
of BPD3 to the slow-counter readout board.
It did not change the crate of residence.
  
  20   Fri, 24.06.2022 19:19:35 B. Maksiak, S. PulawskiBoard 136 moved. TOF -> BPD2The board 136 was moved from mRPC (TOF1 crate)
to BPD2
  
  19   Thu, 23.06.2022 11:57:37 B. Maksiak, B. RumbergerBoards 142, 143 moved. PSD -> FTOFThe boards 142 and 143 were moved from PSD
to FTOF.
  
  18   Sat, 21.05.2022 20:54:59 B. MaksiakController SN5 has issues with programming via JTAGIt cannot program even one board due to many
strange errors.
  
  17   Sat, 21.05.2022 17:47:18 B. MaksiakProgramming boards via JTAG on the controllerProgramming boards using USB-Blaster connected
to the controller in the crate is painful.
One cannot program more than two boards at
  
  16   Sat, 21.05.2022 17:44:27 B. MaksiakIssue with controller SN6The controller SN6 does not provide properly
the sample clock. When inserted, the PLL-not-locked
diodes in the same crate blink in a random
  
  15   Sat, 21.05.2022 17:41:36 B. MaksiakStrange behavior of SN156The board SN156 behaves in a strange way.
When it is put to the crate with other boards
then the board overrides the readout of all
  
  14   Fri, 29.04.2022 10:55:12 B. MaksiakProblems with board x106 (SN154)The board SN154 shows strange readouts on
the first chip. It needs to be exchanged
with other board.
 x106_problem.pdf 
  13   Fri, 25.02.2022 12:30:27 B. Maksiak, Y. BalkovaPlans for March and April periods[B]To do before shifts in March[/B]

[LIST]
  
  12   Mon, 6.09.2021 17:28:18 B. MaksiakPreparation of DRS readout machinesAll DRS readout machines were installed and
prepared. All of them are Dell blades with
24 GB of RAM and USB PCI Express expansion
  
  11   Sun, 5.09.2021 10:16:01 B. MaksiakStatus of PCI Express USB 3.0 cards[TABLE border="1"]
No. | Status | Location | Info
|-
  
  10   Mon, 30.08.2021 14:24:56 B. MaksiakBPD1A machine connected to TDAQ networkTwo days ago the first BPD1 readout machine
was connected to TDAQ network. The network
cable labeled "[B]DRSBPD1 01[/B]" was connected
  
  9   Fri, 27.08.2021 21:00:20 B. MaksiakNetwork cables to BPDsYesterday and today the network cables from
DAQ room to BPDs were pulled.
  
  8   Fri, 23.07.2021 11:00:18 B. MaksiakFirst readout from BPD1Yesterday we connected one DRS board to BPD1
and checked the basic connectivity and readout.
Due to delay in delivery of 750 meters of
  
  7   Tue, 20.07.2021 23:08:29 B. MaksiakTODO[B]Problems[/B]
[LIST]
[*] Wiener crates have currents and voltages
  
  6   Mon, 19.07.2021 09:42:09 Y. Balkova, B. MaksiakInstruction for test data taking[LINE]
[B]ssh to supervisor machine and all machines,
which are used for the data taking[/B]
  
  5   Sat, 17.07.2021 10:46:17 B. MaksiakDRS config registersRegister 13, trigger register
0b
IGNORE DISCARD - to work without backplane
  
  4   Tue, 13.07.2021 16:10:00 Y. Balkova, B. MaksiakRotary switches map[LINE]
Rotary switches of each DRS board have XXX
format, where X can be a letter from A to
  
  3   Mon, 12.07.2021 09:39:14 Y. Balkova, B. MaksiakInstruction for DRS boards[LINE]
[B]Program FPGA[/B]
  
  2   Mon, 12.07.2021 09:22:41 Y. Balkova, B. MaksiakInstruction for DRS machines[B]Register the machine in CERN network[/B]
MAC address --> add the machine to CERN network:
network.cern.ch
  
  1   Sat, 10.07.2021 12:17:37 B. Maksiak, Y. BalkovaAdventures with DRS machinesThe first machine that we tried to set up
was an ex-ALICE HP ProLiant DL160 G6. After
the installation of PCI-E USB 3.0 card, it
  
ELOG V3.1.3-7933898